Stable high current calibrator with 120 A AC/DC range, built-in multimeter, floating outputs and both real and simulated transconductance amplifier functions.
- High accuracy throughout all ranges
- 8 Vpk floating outputs
- Real and simulated transconductance amplifiers
Calibration domain
The main purpose of M151 is calibration of ammeters or use as a current amplifier but it can be used as well for transducer calibration thanks to its built-in multimeter, as stable current source for contact resistance measurement or as either real or simulated transconductance amplifier.
RS-232 and GPIB interface allow for remote control of the calibrator and time saving automated calibrations using SW package CALIBER/WinQbase.
DC current | 8 mA - 120 A (3000 A with 151-25 Current Coil) ± 250 ppm / year |
AC current | 8 mA - 120 A, 15 Hz - 1 kHz (3000 A with 151-25 Current Coil) ± 250 ppm / year |
Process Multimeter
DC voltage | 0 - 20 V ± 100 ppm / year |
AC voltage | 1 - 20 V ± 200 ppm / year |
DC current | 0 - 200 mA ± 100 ppm / year |
AC current | 10 - 200 mA ± 200 ppm / year |
Frequency | 1 Hz - 10 kHz ± 50 ppm / year |